This is an aftermarket part, not an original manufacturer part. Instrument names and part numbers identify what it fits and replaces; they remain the trademarks of their owners. The manufacturer, country of origin, rated life and warranty are stated in the quote, before you decide anything. If you need the OEM part, say so and I will quote that instead.

Pump and injector spares

Sapphire plunger rod for Agilent 1220, 1260 and 1290

Replaces OEM part number: 5067-4695

Sapphire plunger rod for the Agilent 1220, 1260 and 1290, cross-referenced to 5067-4695. These are the later pump heads, and on a 1290 running near its pressure ceiling the plunger and seal both wear faster than the same parts in an 1100 ever did, so the service interval that suited your old stack will not suit this one. Inspect the rod when the seal is changed. A scored band where the seal rides means both parts go together, whatever the hours say.

Type
Sapphire plunger rod
Fits
Agilent 1220, 1260, 1290
Replaces
5067-4695
Condition
new
Price
Quoted per order, in Canadian dollars

Before you order: if the instrument is misbehaving, tell me the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, I will say so.