Skip to main content

Independent. Diagnosis first, parts second.

+1 647-919-3041 [email protected]

Chromkit
Get a quote
Illustration of an injection valve rotor seal. Generic drawing, not a photograph of this part. Illustration

This is an aftermarket part, not an original manufacturer part. Instrument names and part numbers identify what it fits and replaces; they remain the trademarks of their owners. The manufacturer, country of origin, rated life and warranty are stated in the quote, before you decide anything. If you need the OEM part, say so and I will quote that instead.

Pump and injector spares

Rotor seal, Vespel, 2-groove, for Agilent 1100 and 1200

Replaces OEM part number: 0100-1853

Two groove Vespel rotor seal for the Agilent 1100 and 1200 injection valve, cross-referenced to 0100-1853. Count the grooves on the seal you take out before ordering, because the two groove and the standard pattern are not interchangeable and the valve body will accept the wrong one physically. Vespel is a polyimide, so it tolerates most solvents and dislikes high pH: mobile phases above about pH 10 shorten its life noticeably. A scored seal shows up as carryover first and injection precision second.

Type
Rotor seal, Vespel, 2-groove
Fits
Agilent 1100, 1200
Replaces
0100-1853
Condition
new
Price
Quoted per order, in Canadian dollars

Before you order: if the instrument is misbehaving, tell me the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, I will say so.

How buying works, written for procurement →