Skip to main content

+1 437-505-5995 [email protected]

Waters ACQUITY UPLC stack: PDA detector, Column Manager, Sample Manager and Binary Solvent Manager. refurbished

A refurbished system, serviced and performance-checked before it was listed, held in stock in the Greater Toronto Area. The configuration, software licence and warranty are written into the quote before you decide anything. You buy it from Chromkit and we invoice you, so your contract and your warranty are with us. Installation, a service contract and parts are available too, quoted separately, so one supplier looks after the instrument once it is on your bench.

HPLC and UHPLC

Waters ACQUITY UPLC with PDA detector and Column Manager, refurbished

Price on request refurbished

Type
UPLC (UHPLC), binary gradient
Pump
Binary Solvent Manager, high-pressure mixing of two solvents chosen from four (A1/A2, B1/B2), built-in six-channel vacuum degasser, 0.001 to 2.000 mL/min, up to 15,000 psi (1034 bar) at 1 mL/min
Autosampler
Sample Manager, 10 µL loop as standard (1 to 50 µL loops available), sample compartment cooled to 4 °C, 2 mL vial or microplate trays
Column compartment
Column Manager, holds two columns up to 150 mm (four up to 50 mm), 4 to 90 °C, two switching valves for column selection
Detector
ACQUITY UPLC PDA (photodiode array), 190 to 500 nm, records the full UV spectrum across every peak
Detector specification
Waters published figures for the detector: wavelength accuracy ±1 nm, noise 14 µAU with the analytical flow cell, linearity within 5% at 2.0 AU, data rate 1 to 80 Hz
Flow cell
Rated to 69 bar maximum. Flush and store it as the Waters PDA guide describes, and pre-flush a new column before connecting it to the cell
Software
Waters Empower 3, licence transfers with the system. PC not included
Location
Mississauga, Ontario
Condition
refurbished
Price
Quoted per order, in Canadian dollars

About this instrument

A four-module ACQUITY stack built for method development: binary pump, cooled sample manager, a Column Manager that switches between columns, and a photodiode array detector. The photograph is the actual unit. The PDA is the reason to choose this system over a single-wavelength one. It records the full spectrum from 190 to 500 nm across every peak, so you can choose the wavelength after the run, check peak purity, and compare a peak's spectrum against a reference standard. Peak purity is the usual way to show specificity in a stability-indicating method. Two things to know for spectral work. Keep the peak below 1 AU, because above that the photometric error changes the shape of the spectrum, and Waters says so in its PDA guide. And compare spectra taken at the same mobile phase composition, because pH and solvent change the spectrum too. The Column Manager holds two columns up to 150 mm, or four up to 50 mm, so you can screen columns without re-plumbing. Like any UPLC, it will run your HPLC methods only after checking. The system volume is much smaller, so a gradient method elutes earlier, and moving to smaller particle columns is a change you have to justify under USP <621>. The pump is binary, so methods that mix three or four solvents are out. Tell us what you need to run and we will tell you whether this fits.

Before you order: if the instrument is misbehaving, tell us the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, we will say so.

How buying works, written for procurement →

Parts we list for Waters systems

Lamps, seals and valves wear out on every system. Check each part's "Fits" line against your modules, or send us the module numbers and we will match them.

All 43 for Waters →

Waters ACQUITY UPLC with PDA detector and Column Manager, refurbished

Price on request

Request a quote