Illustration
This is an aftermarket part, not an original manufacturer part. Instrument names and part
numbers identify what it fits and replaces; they remain the trademarks of their owners. The
manufacturer, country of origin, rated life and warranty are stated in the quote, before you
decide anything. If you need the OEM part, say so and I will quote that instead.
Pump and injector spares Manual injection valve rotor seal for Agilent 1100, 1200 and Rheodyne 7125
Replaces OEM part numbers:
Agilent 0101-0623, PerkinElmer 09904802, Rheodyne 7125-047
Aftermarket manual injection valve rotor seal for the Agilent 1100, 1200, PerkinElmer 200 Series, Rheodyne 7125, see the full fitment list above, cross-referenced to Agilent 0101-0623. A worn seal leaks between ports in the injection valve, which shows up as carryover, as low injection volume, or as a pressure step at the moment of injection.
- Type
- Rotor seal, Vespel, manual injection valve
- Fits
- Agilent 1100, 1200. PerkinElmer 200 Series. Rheodyne 7125, 7126, 7725, 7725i, 9725
- Replaces
- Agilent 0101-0623, PerkinElmer 09904802, Rheodyne 7125-047
- Condition
- new
- Price
- Quoted per order, in Canadian dollars
Before you order: if the instrument is misbehaving, tell me the symptom, not the part
number. Lamps and seals get replaced for problems they did not cause. The first look costs
nothing, and if this part is not the fix, I will say so.
How buying works, written for procurement →