Skip to main content

Independent. Diagnosis first, parts second.

+1 647-919-3041 [email protected]

Chromkit
Get a quote
Illustration of an injection needle and needle seat. Generic drawing, not a photograph of this part. Illustration

This is an aftermarket part, not an original manufacturer part. Instrument names and part numbers identify what it fits and replaces; they remain the trademarks of their owners. The manufacturer, country of origin, rated life and warranty are stated in the quote, before you decide anything. If you need the OEM part, say so and I will quote that instead.

Pump and injector spares

Injection needle for Shimadzu LC-2010

Replaces OEM part number: 228-41024-91

Injection needle for the LC-2010, uncoated, 10 Series, cross-referenced to 228-41024-91. Uncoated is worth noting, because a coated needle is the usual answer to sample carryover with sticky or basic compounds, and this is not that part. A blunted or bent tip damages the needle seal below it, so if the needle has done a lot of injections, plan to change both. Carryover that a stronger wash does not fix, on a needle that has been in for years, usually means the tip rather than the wash.

Type
Injection needle, uncoated, 10 Series
Fits
Shimadzu LC-2010
Replaces
228-41024-91
Condition
new
Price
Quoted per order, in Canadian dollars

Before you order: if the instrument is misbehaving, tell me the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, I will say so.

How buying works, written for procurement →

More for Shimadzu and similar systems