Illustration
Spectrophotometer and AAS lamps
Lamp for Beckman DU730
Lamp for the Beckman DU730, cross-referenced to A23778. The DU730 is a routine life science bench, used mostly for nucleic acid and protein quantitation at fixed wavelengths rather than for full spectra. That matters, because a lamp can be well past its best at 260 nm while a 595 nm assay still reads normally, and nobody notices until the A260/A280 ratios start drifting. If your ratios have moved and your samples have not, read the lamp hours before you re-prepare anything.
- Type
- Detector lamp
- Fits
- Beckman DU730
- Replaces
- A23792
Price on request