This is an aftermarket part, not an original manufacturer part. Instrument names and part numbers identify what it fits and replaces; they remain the trademarks of their owners. The manufacturer, country of origin, rated life and warranty are stated in the quote, before you decide anything. If you need the OEM part, say so and I will quote that instead.

Spectrophotometer and AAS lamps

Lamp for iCE3 cIEF analyser

Replaces OEM part number: 102-235

Lamp for the iCE3 cIEF analyser, cross-referenced to 102-235. Whole column imaging detection reads absorbance along the entire capillary at once rather than at a single point, so a lamp with uneven or falling output shows up as a baseline that slopes across the pI range rather than as general noise. Peaks at one end of the focus look wrong while the other end looks fine. If your pI markers have shifted relative to each other rather than together, the source is worth checking before the capillary.

Type
Detector lamp
Fits
ProteinSimple iCE3 cIEF
Replaces
102-235
Condition
new
Price
Quoted per order, in Canadian dollars

Before you order: if the instrument is misbehaving, tell me the symptom, not the part number. Lamps and seals get replaced for problems they did not cause. The first look costs nothing, and if this part is not the fix, I will say so.